    Bench-TOP pH Meter, Model : HD2205.2 - HD2206.2 - HD2256.2 - HD2259.2 - HD22569.2, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    Introduction
    The instrument series HD2205.2, HD2206.2, HD2256.2, HD2259.2, and
    HD2569.2 is made up of bench top instruments for electrochemical measures:
    pH, conductivity, dissolved oxygen, and temperature. They are fitted with a large backlighted LCD display.
    The HD2205.2 is equipped with two BNC inputs for the measurement of pH, mV, redox potential ( ORP) with pH or redox electrodes, or electrodes with separate
    reference, and one input for combined pH/ temperature probes fitted with SICRAM module.
    The HD2206.2 measures conductivity, resistivity in liquids, total disssolved solids ( TDS) , and salinity with combined 4-ring and 2-ring conductivity/ temperature
    probes. The conductivity probes can have a direct input or with SICRAM module. The inputs are separate.
    The HD2256.2 measures pH, mV, redox potential ( ORP) with pH, redox electrodes or electrodes with separate reference. Conductivity and resistivity in liquids, total dissolevd solids ( TDS) and salinity with combined 4-ring and 2-ring conductivity/ temperature probes. The conductivity probes can have a direct input
    or with SICRAM module. The inputs are separate.
    The HD2259.2 measures pH, mV, redox potential ( ORP) with pH, redox electrodes or electrodes with separate reference; the concentration of dissolved oxygen in liquids ( in mg/ l) , and saturation index ( in % ) , using SICRAM combined
    probes of polarographic type with two or three electrodes and integrated temperature
    sensor.
    The HD22569.2 measures pH, mV, redox potential ( ORP) with pH, redox electrodes or electrodes with separate reference; conductivity, resistivity in liquids, total dissolved solids ( TDS) and salinity with combined 4-ring and 2-ring conductivity/ temperature probes with direct input or SICRAM module; concentration of dissolved oxygen in liquids ( in mg/ l) and saturation index ( in % ) , using
    SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor.
    Al models are fitted with input for the measurement of temperature with Pt100 or Pt1000 immersion, penetration or contact probes. The probes are equipped with an automatic detection module, with the factory calibration settings already being memorized inside.
    The pH electrode calibration can be carried out on one or five points and the calibration sequence can be chosen from a list of 13 buffers.
    • The pH electrode calibration can be carried out on one or five points and the calibration sequence can be chosen from a list of 13 buffers Temperature compensation can be automatic or manual.
    • The conductivity probe calibration can be performed automatically with automatically
    detected conductivity calibration solutions: 147¼ S/ cm, 1413¼ S/ cm, 12880¼ S/ cm or 111800¼ S/ cm or manually with calibration solutions having different
    values.
    • The dissolved Oxygen probe' s quick calibration function guarantees timely correctness of the performed measurements.
    • Conductivity, dissolved oxygen and temperature probes fitted with SICRAM
    module can store factory and calibration data inside.
    The instruments of the series HD22… are datalogger, they memorize up to 2, 000 samples of data:
    • pH, mV and temperature: HD2205.2
    • conductivity or resistivity or TDS or salinity and temperature: HD2206.2
    • pH or mV, conductivity or resistivity or TDS or salinity and temperature: HD2256.2,
    • pH or mV, concentration of dissolved oxygen or saturation index and saturation index and temperature: HD2259.2,
    • pH or mV, conductivity or resistivity or TDS or salinity, concentration of dissolved
    oxygen and temperature: HD22569.2.
    The data can be transferred from the instrument connected to a PC via the multistandard
    RS232C serial port and USB 2.0. The storing parameters can be configured using the menu.
    The RS232C serial port can be used to transfer the acquired measurements to a 24 column portable printer in real time ( S’ print-BT) .
    The instruments equipped with HD22BT ( Bluetooth) option can transfer data without any connection to a PC or printer fitted with Bluetooth input or through Bluetooth/ RS232C converter.
    The software DeltaLog11 allows instrument management and configuration, and data processing on PC.
    The instruments have IP66 protection degree.

    Technical characteristics of the instruments series HD22…
    Common technical data
    Instrument
    Dimensions ( Length x Width x Height) 265x185x70mm
    Weight 490g
    Materials ABS, rubber
    Display Back lighted, matrix point display.
    240x64 points, visible area: 128x35mm
    Operating conditions
    Working temperature -5 … 50° C
    Storing temperature -25 … 65° C
    Working relative humidity 0 … 90% R.H. without condensate
    Protection degree IP66
    Power
    Mains adapter ( cod. SWD10)
    12Vdc/ 1A
    Auxiliary socket For supplying of electrode holder with built-in
    stirrer HD22.2
    Security of memorized data
    Unlimited
    Time
    Date and hour Real time schedule with backup battery E
    3.6V - ½ AA
    Accuracy 1min/ month max drift
    Measured values storing
    Quantity 2000 screens
    Storage interval 1s … 999s
    Calibration storage
    Quantity Last 8 calibrations of each physical
    quantity
    RS232C serial interface
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 115200 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Length of serial cable Max 15m
    USB Interface
    Type 1.1 - 2.0 electrically isolated
    USB Interface optional
    EMC standard regulations
    Security EN61000-4-2, EN61010-1 level 3
    Electrostatic discharge EN61000-4-2 level 3
    Electric fast transients EN61000-4-4 level 3,
    EN61000-4-5 level 3
    Voltage variations EN61000-4-11
    Electromagnetic interference
    susceptibility IEC1000-4-3
    Electromagnetic interference
    emission EN55020 class B

    Technical characteristics HD2205.2 pH - mV - ° C - ° F measurement
    Measured values
    pH - mV - ° C - ° F
    Connections
    Input for temperature probes 8-pole male DIN45326 connector with SICRAM modulen
    Inputs pH/ mV j - k female BNC
    Inputs for SICRAM module 8-pole male DIN45326 connector
    pH/ temperature l a. m
    Serial interface DB9 connector ( 9- pole male)
    USB interface USB connector type B
    Bluetooth Optional
    Mains adapter 2- pole connector ( Ø 5.5mm-2.1mm) .
    Positive at centre.
    Socket for power supply of electrode 2 -pole connector ( Ø 5.5mm-2.1mm) .
    holder with built-in magnetic stirrer Positive at centre ( output 12Vdc/ 200mA max) .
    Measurement of pH by instrument
    Measuring range -9.999… + 19.999pH
    Resolution 0.01 o 0.001pH selectable from menu
    Accuracy ± 0.001pH ± 1digit
    Input impedance > 1012©
    Calibration error @ 25° C | Offset| > 20mV
    Slope > 63mV/ pH o Slope 106.5% or Senstivity < 85%
    Calibration points Up to 5 points from a list of 13 automatically detected buffers.
    Automatically detected pH standard 1.679pH - 2.000pH - 4.000pH - 4.008pH - 4.010pH
    solutions ( @ 25° C) 6.860pH - 6.865pH - 7.000pH - 7.413pH - 7.648pH - 9.180pH - 9.210pH - 10.010pH
    Measurement of mV by instrument
    Measuring range -1999.9… + 1999.9mV
    Resolution 0.1mV
    Accuracy ± 0.1mV ± 1digit
    Drift after 1 year 0.5mV/ year
    Measurement of temperature by instrument
    Pt100 Measuring range -50… + 150° C
    Pt1000 Measuring range -50… + 150° C
    Resolution 0.1° C
    Accuracy ± 0.1° C ± 1digit
    Drift after 1 year 0.1° C/ year