    CONDUCTIVITY METER - THERMOMETER, Model : HD2306.0, Brand : DeltaOhm

    [29 Nov. 2010, 3:35:29]
    CONDUCTIVITY METER - THERMOMETER,  Model : HD2306.0,  Brand : DeltaOhm
    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    The HD2306.0 is a portable instrument with a large LCD display. It measures conductivity, liquid resistivity, total dissolved solids ( TDS) , and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature only is
    measured by Pt100 or Pt1000 immersion, penetration or contact probes.
    The probe calibration can be performed automatically in one or more than one of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions. The temperature probe is fitted with an automatic detection module, with the factory calibration settings already being memorized inside. The Max, Min and Avg function calculates the maximum, minimum or average values. Other functions include: the relative measurement REL, the Auto-HOLD function, and the automatic turning off which can also be disabled.
    The instrument has IP67 protection degree.

    INSTRUMENT TECHNICAL CHARACTERISTICS
    Instrument
    Dimensions
    ( Length x Width x Height) 140x88x38mm
    Weight 160g ( complete with batteries)
    Materials ABS
    Display 2x41D2 digits plus symbols
    Visible area: 52x42mm
    CONDUCTIVITY METER - THERMOMETER HD2306.0
    Operating conditions
    Working temperature -5… 50° C
    Storing temperature -25… 65° C
    Working relative humidity 0… 90% RH without condensation
    Protection degree IP67
    Power
    Batteries 3 1.5V type AA batteries
    Autonomy 200 hours with 1800mAh alkaline batteries
    Power absorbed with instrument off 20¼ A
    Connections
    Conductivity input/ temperature probes 8-pole male DIN45326 connector
    Measurement of conductivity by Instrument
    Resolution with K cell= 0.1 0.01¼ S in range 0.00… 19.99¼ S
    Measurement range
    ( K cell= 1) / Resolution 0.0… 199.9¼ S / 0.1¼ S
    200… 1999¼ S / 1¼ S
    2.00… 19.99mS / 0.01mS
    20.0… 199.9mS / 0.1mS
    Accuracy ( conductivity) ± 0.5% ± 1digit
    Measurement of resistivity by Instrument
    Measurement range / Resolution 4.0… 199.9© / 0.1©
    200… 999© / 1©
    1.00k… 19.99k© / 0.01k©
    20.0k… 99.9k© / 0.1k©
    100k… 999k© / 1k©
    1… 10M© / 1M©
    Accuracy ( resistivity) ± 0.5% ± 1digit
    Measurement of total dissolved solids ( with coeffi cient Ç / TDS= 0.5)
    Resolution with K cell= 0.1 0.05mg/ l in range 0.00… 19.99mg/ l
    Measurement range
    ( K cell= 1) / Resolution 0.0… 199.9 mg/ l / 0.5 mg/ l
    200… 1999 mg/ l / 1 mg/ l
    2.00… 19.99 g/ l / 0.01 g/ l
    20.0… 199.9 g/ l / 0.1 g/ l
    Accuracy ( total dissolved solids) ± 0.5% ± 1digit
    Measurement of temperature by Instrument
    Pt100 measurement range -50… + 200° C
    Pt1000 measurement range -50… + 200° C
    Resolution 0.1° C
    Accuracy ± 0.25° C
    Drift after 1 year 0.1° C/ year
    Automatic/ manual temperature
    compensation 0… 100° C with ± T= 0.00… 4.00% / ° C
    Reference temperature 20° C or 25° C
    Ç / TDS conversion factor 0.4… 0.8
    Cell constant K ( cm-1) 0.1, 0.7, 1.0 and 10.0
    Standard solutions
    automatically detected ( @ 25° C)
    147 ¼ S/ cm
    1413 ¼ S/ cm
    12880 ¼ S/ cm
    111800 ¼ S/ cm