Selamat Datang di Indotrade.co.id
Masuk | Daftar

Selamat Datang di

MIKASE LABORATORY - PT MITRA KATIGA SEJAHTERA / SUPPLIER ALAT UJI KUALITAS UDARA, AIR, CUACA , KEBISINGAN SERTA GETARAN

http://indotrade.co.id/mikaselaboratory/

Anggota Prioritas Prioritas Indotrade
Tahun ke-16

menampilkan 64 dari 233
  • Katalog Produk

    pH METER - CONDUCTIVITY METER - THERMOMETER, Model : HD2156.1 AND HD2156.2, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    The HD2156.1 and HD2156.2 are portable instruments with a large LCD display. They measure pH, mV, redox potential ( ORP) , conductivity, liquid resistivity, total dissolved solids ( TDS) and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature only is measured by Pt100 or Pt1000 immersion, penetration or contact probes.
    The pH electrode calibration, as well as manual, can be carried out on one, two or three points and the calibration sequence can be chosen from a list of 13 buffers. The probe calibration can be performed automatically in one or more of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions. The HD2156.2 instrument is a datalogger. It memorizes up to 20, 000 sets of
    three measurements composed of pH or mV, conductivity or resistivity or TDS or salinity and temperature: these data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The
    storing interval, printing, and baud rate can be configured using the menu. The HD2156.1 and HD2156.2 models are fitted with an RS232C serial port and can transfer the acquired measurements to a PC or to a portable printer in real time.
    The Max, Min and Avg function calculates the maximum, minimum or average values. Other functions include: the Auto-HOLD function and the automatic turning off which can also be disabled.
    The instruments have IP67 protection degree.

    INSTRUMENT TECHNICAL CHARACTERISTICS
    Instrument
    Dimensions
    ( Length x Width x Height) 185x90x40mm
    Weight 470g ( complete with batteries)
    Materials ABS, rubber
    Display 2x41D2 digits plus symbols
    Visible area: 52x42mm
    Operating conditions
    Working temperature -5… 50° C
    Storing temperature -25… 65° C
    Working relative humidity 0… 90% RH without condensation
    Protection degree IP67
    Power
    Batteries 4 1.5V type AA batteries
    Autonomy 200 hours with 1800mAh alkaline batteries
    Power absorbed with instrument off 20¼ A
    Mains Output mains adapter 9Vdc / 250mA
    Security of memorized data Unlimited, independent of battery charge conditions
    Time
    Date and time Schedule in real time
    Accuracy 1min/ month max error
    Measured values storage - model HD2156.2
    Type 2000 pages containing 10 samples each
    Quantity 20, 000 sets of three measurements composed
    of pH or mV, Ç or © or TDS or salinity and temperature.
    Storage interval 1s… 3600s ( 1hour)
    Serial interface RS232C
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 38400 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Serial cable length Max 15m
    Immediate print interval 1s… 3600s ( 1hour)
    USB interface - model HD2156.2
    Type 1.1 - 2.0 electrically isolated
    Connections
    pH/ mV input Female BNC connector
    Conductivity input 8-pole male DIN45326 connector
    Serial interface and USB 8-pole MiniDin connector
    Mains adapter 2-pole connector ( positive at centre)
    Measurement of pH by Instrument
    Measurement range -2.000… + 19.999pH
    Resolution 0.01 or 0.001pH selectable from menu
    Accuracy ± 0.001pH ± 1digit
    Input impedance > 1012©
    Calibration error @ 25° C | Offset| > 20mV
    Slope > 63mV/ pH or Slope 106.5% or Sensitivity < 85%
    Measurement of mV by Instrument
    Measurement range -1999.9… + 1999.9mV
    Resolution 0.1mV
    Accuracy ± 0.1mV ± 1digit
    Drift after 1 year 0.5mV/ year
    Measurement of conductivity by Instrument
    Resolution with K cell= 0.1 0.01¼ S/ cm in range 0.00… 19.99¼ S/ cm
    Measurement range ( K cell= 1)
    Resolution 0.0… 199.9¼ S/ cm / 0.1¼ S/ cm
    200… 1999¼ S/ cm / 1¼ S/ cm
    2.00… 19.99mS/ cm / 0.01mS/ cm
    20.0… 199.9mS/ cm / 0.1mS/ cm
    Accuracy ( conductivity) ± 0.5% ± 1digit
    Measurement of resistivity by Instrument
    Measurement range / Resolution 4.0… 199.9© / 0.1©
    200… 999© / 1©
    1.00k… 19.99k© / 0.01k©
    20.0k… 99.9k© / 0.1k©
    100k… 999k© / 1k©
    1… 10M© / 1M©
    Accuracy ( resistivity) ± 0.5% ± 1digit
    Measurement of total dissolved solids ( with coefficient Ç / TDS= 0.5)
    Resolution with K cell= 0.1 0.05mg/ l in range 0.00… 19.99mg/ l
    Measurement range
    ( K cell= 1) / Resolution 0.0… 199.9 mg/ l / 0.5 mg/ l
    200… 1999 mg/ l / 1 mg/ l
    2.00… 19.99 g/ l / 0.01 g/ l
    20.0… 199.9 g/ l / 0.1 g/ l
    Accuracy ( total dissolved solids) ± 0.5% ± 1digit
    Measurement of salinity
    Measurement range / Resolution 0.000… 1.999g/ l / 1mg/ l
    2.00… 19.99g/ l / 10mg/ l
    Accuracy ( total dissolved solids) ± 0.5% ± 1digit
    Automatic/ manual temperature
    compensation 0… 100° C with ± T that can be selected from
    0.00 to 4.00% / ° C
    Reference temperature 20° C or 25° C
    Ç / TDS conversion factor 0.4… 0.8
    Cell constant K ( cm-1) 0.1, 0.7, 1.0 and 10.0
    Standard solutions automatically detected ( @ 25° C)
    147 ¼ S/ cm
    1413 ¼ S/ cm
    12880 ¼ S/ cm
    111800 ¼ S/ cm
    Measurement of temperature by Instrument
    Pt100 measurement range -50… + 200° C
    Pt1000 measurement range -50… + 200° C
    Resolution 0.1° C
    Accuracy ± 0.25° C
    Drift after 1 year 0.1° C/ year