Keterangan
The HD2010UC is an integrating portable sound level meter performing statistical analyses. The instrument has been designed combining price competitiveness and simplicity of use. Attention has been paid
to the possibility of updating the instrument. Besides, the HD2010UC can be integrated with other options to extend its application scope at any time. The firmware can be upgraded directly by the user using the DeltaLog5 program supplied with the instrument.
Technical characteristics
Standards Class 1 or 2 group X according to IEC 61672: 2002 and class 1 or 2 according to IEC 60651: 2001 and IEC 60804: 2000
type 1 or 2 according to ANSI S1.4-1983 and S1.43-1997
½ inch Microphone UC52 condenser type, pre-polarized, for free field
Dynamic range 30 dBA ÷ 143 dB Peak
Linear Field 80 dB
Acoustic Parameters Spl, Leq, SEL, LEP, d, Lmax, Lmin, Lpk, Dose, Ln
Frequency Weighting Simultaneous A, C, Z ( only C and Z for Lpk)
Temporal Weighting Simultaneous FAST, SLOW, IMPULSE
Integration From 1s to 99 hours with Back-Erase function
Statistical Analysis It displays up to 3 percentile levels, from L1 to L99
â Data Loggerâ and â Advanced Analyzerâ options
Probability distribution and percentile level calculation from L1 to L99
Parameter: LFp, Leq, Lpk A, C or Z -weighted( only C or Z for Lpk)
Sampling frequency: 8 samples/ second
Classification: Classes of 0.5 dB
Analysis of Events â Data Loggerâ and â Advanced Analyzerâ options
Calculation of 5 freely-programmable event parameters
Calculation of statistical levels from L1 to L99
Event identification trigger with programmable threshold and duration filter
Manual trigger
Profile Data Logging â Data Loggerâ option
1 profile with programmable sampling from 1/ 8 s to 1 hour and 3 profiles with 2 samples/ second
Display Graphic display 128x64
3 parameters in numeric format
â LCDâ option
Backlit LCD
â Data Loggerâ option
Profi le LAFp with 8 samples/ second
â Advanced Analyzerâ option
Graph of sound level probability distribution
Graph of percentile levels from L1 to L99
Memory Internal, equal to 2 MB, enough to store over 500 recordings
â Data Loggerâ option
Internal, equal to 4 MB ( 1 profi le for 23 hours or over 80 recording days of 3 parameters per minute) . Expandable to 8 MB
External, via the HD2110MC memory card interface, using MMC or SD cards up to 2 GB
Input/ Output RS232 serial and USB interfaces
AC output ( LINE)
DC output
PC Programs DeltaLog5: PC interface for data download, setup and sound level meter management ( supplied with the instrument)
DL5 Monitor: For real-time acquisition in the PC mass storage, scheduler, audio recording
DL5 Noise Studio: Analysis modular program
â Worker Protectionâ : Analysis module according to decree 195/ 2006
â Railway Traffi câ : Analysis module of train noise profiles according to the decree of 16/ 03/ 1998
Operating conditions Working temperature -10÷ 50° C, 25÷ 90% RH ( without condensation) , 65÷ 108kPa. Protection degree: IP64
Power ⢠4 alkaline or rechargeable NiMH type AA batteries or external 9÷ 12Vdc 300mA
Dimension and weight ⢠445x100x50 mm equipped with preamplifier, 740 g ( with batteries)