    HD2003 and HD2003.1 are three axis ultrasonic anemometers, they measure
    the speed and direction of wind, the U-V-W Cartesian components of speed, sound speed and sonic temperature.
    The HD2003 allows also to detect temperature and relative humidity of the air and barometric pressure.
    The HD2003 main features are:
    • Determination of the anemometric quantities represented in diverse measurement
    units: wind speed and direction, U-V-W Cartesian components of speed, sound speed, sonic temperature.
    • ( HD2003 Model) additional output quantities: Temperature, Relative Humidity
    and Pressure.
    • 5 analogue voltage or current outputs, with different measuring ranges.
    • RS232 and Multidrop RS485 Serial Communication interfaces.
    • Configurable output rate of digital output data string.
    • Configurable average periods 1÷ 60sec and 1÷ 60min. for all output quantities.
    • Algorithmic raw date processing and validation, assuring ± 1% precision to
    anemometric quantities.
    • Digital high frequency data acquisition mode with 50Hz data output.
    • Self-diagnosis with error checking and report.
    • Reliability and precision on whole measuring range, no additional calibration required.
    • Flexible, easy-to use operating software, confi gurable according to the user’ s
    needs through Computer interface.
    • User interface for ‘ Setup’ management and software upgrade through RS232 or
    RS485.
    • Automatic alignment to the magnetic North through built in compass.
    • No moving part, with reduced maintenance and service costs.
    • Rugged and reliable structure, suitable for continuous operation even in severe environmental conditions.
    • Low power consumption.
    • ( On request) Heaters Option: built-in heating device of sonic transducers, to prevent ice and snow formation. Assures correct measurements even in presence of sleet or snow.
    Typical applications:
    - Meteorology
    - Aviation and Navigation
    - Tunnels, Highways
    - Climatology
    - Sport and winter stations
    - Safety in yards
    - Industrial buildings
    Technical specifications
    Output quantities
    • Anemometric parameters Wind speed and direction, Sound Speed, Sonic Temperature, U-V-W Components
    • Meteorological parameters ( Model HD2003) Pressure, Temperature, Relative Humidity
    • Heading Compass with magnetic Azimuth
    • Moving Averages 1÷ 60 sec./ 1 ÷ 60 min.
    • Output rate 1÷ 50Hz ( RS232 or RS485)
    Wind Speed
    • Measuring unit m/ s, cm/ s, km/ h, knots, mph
    • Range 0÷ 60 m/ s ( 216 km/ h)
    • Resolution 0.01 m/ s
    • Accuracy ± 1% of reading
    Wind Direction
    • Range Azimuth: 0÷ 360° Elevation: ± 60°
    • Resolution 0. 1°
    • Accuracy ± 1°
    Sound speed
    • Range 300 ÷ 380 m/ s
    • Resolution 0.01 m/ s
    • Accuracy ± 1% of reading
    Sonic Temperature
    • Range -40 + 60° C
    • Resolution 0.1 ° C
    • Accuracy ± 1° C
    Compass
    • Range 0 ÷ 360°
    • Resolution 0.1 °
    • Accuracy ± 1°
    Digital Outputs
    • Communications RS-232 full duplex, Multidrop RS-485 half duplex
    • Baud Rate 9600 ÷ 115200 bit/ sec.
    • Output Rate Normal functioning mode: 1 ÷ 3600 sec
    Digital high frequency: 1/ 50 sec
    • Measured data Digital string of anemometric quantities and compass ( Model HD2003) Pressure, temperature, relative
    humidity
    Relative Humidity
    Capacitive sensor
    Analog output ( 0 ÷ 100% RH) : 0÷ 20mA, 4÷ 20mA, 0÷ 1V, 0÷ 5V, 1÷ 5V, 0÷ 10V
    Range: 0 ÷ 100% RH
    Resolution 0.1 % RH
    Accuracy ± 2.5% RH @ 23° C
    Analog Outputs
    • Number 5, selectable between all available output quantities
    • Range 0÷ 20mA, 4÷ 20mA, 0÷ 1V, 0÷ 5V, 1÷ 5V, 0÷ 10V
    • Resolution 14 bit max
    Power supply
    • Range 12 ÷ 30 VDC
    • Power