    Relative Pressure Transmitters, Model : HD3604T - HD36V4T, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    The HD3604T... and HD36V4T... are pressure transmitters with microprocessor and current ( 4...20mA) or voltage ( 0...5V, 1...5V or 0...10V) output, respectively.
    The sensor, piezoresistive, is insulated and allows gas and liquid pressure measurement over a wide range of temperatures.
    The measured pressure is relative to atmosphere. The different models cover the scales from 6 to 600bar.
    The case is in stainless steel diam. 20mm and encloses sensor and electronics: in order to connect it to a pressurized system, it is supplied with a ¼ ” BSP threaded connection with a 22mm hex fastening ring.
    For electrical connections, three different solutions are provided:
    • A male connector DIN 43650A,
    • A male connector DIN 43650C,
    • A circular male connector DIN 41524.
    The connectors are fitted with a three or four pole free female socket with fairlead ( according to the models) .
    All transmitters are factory calibrated at three points. The use of a microprocessor circuit allows memorizing the sensor response curve and correcting
    any possible non linearity.
    No calibration is required by the user.

    Technical characteristics Notes
    Output signal
    4 … 20mA HD3604TxBG models
    0 … 5V HD36V4TxBG1 models
    1 … 5V HD36V4TxBG2 models
    0 … 10V HD36V4TxBG3 models
    Bottom scale pressure
    range 6, 10, 16, 25, 40, 60, 100, 160, 250, 400 and 600 bar relative
    Overpressure limit Twice the nominal value
    Sensor Piezoresistive
    Surface contacting the measurement fluid
    Steel 17-4PH
    Fluid contacting the membrane
    Gas or liquid
    Operating temperature -40 … + 125° C
    Power supply voltage
    10… 30Vdc
    15… 30Vdc
    HD36V4TxBG3 models with 0… 10Vdc output
    Absorption < 4mA
    In the models with voltage
    output
    Accuracy d ± 0.25% F.S.
    Temperature effects ± 1% span ± 1% offset For a variation of 100° C
    Mechanical stability