    RH AND TEMPERATURE TRANSMITTERS & MULTIPLATE RADIATION SHIELD, Model : HD9008TR - HD9009TR - HD9007, Brand : DeltaOhm

    [2 Nov. 2010, 20:33:12]
    RH AND TEMPERATURE TRANSMITTERS & MULTIPLATE RADIATION SHIELD,  Model : HD9008TR - HD9009TR - HD9007,  Brand : DeltaOhm
    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    CHARACTERISTICS
    The HD9008TR and HD9009TR are single block RH and temperature microprocessor transmitters, temperature configurable.
    The HD9008TR is a passive transmitter with a 4… 20mA output and 7… 30Vdc power
    supply; the HD9009TR is a transmitter with a 0… 1V standard voltage output ( other outputs available on demand) and 7… 30Vdc power supply.
    Sensors are mounted at the end of a plastic tube: a capacitive humidity sensor and a Platinum temperature sensor ( 100© @ 0° C) .
    The instrument can be reprogrammed by means of a key, and no jumper or potentiometer actions are required. The humidity input can be recalibrated by using two saturated solutions: the fi rst one at 75% , the second one
    at 33% ; the 0% RH… 100% RH relative humidity range is fi xed, 4mA ( or 0Vdc) correspond to 0% RH, 20mA ( or 1Vdc) equal 100% RH.
    Temperature standard confi guration is – 40… + 80° C for the HD9008TR for the HD9009TR, corresponding to 4… 20mA and 0… 1Vdc, respectively.
    The user can configure the temperature output in ranges different from the standard one by means of a Pt100 simulator or of a set of fixed resistances, provided that it is included in the -40° C… + 80° C range with a minimum amplitude of 25° C. Two LEDs give alarm indications ( temperature exceeding set range, sensor breakage or short-circuit) and
    help the operator when programming.
    An out-of-standard temperature operating range can be requested when placing the order.
    Important Warning: probes work in the -40° C… + 80° C temperature range.
    Outside this range data are not correct; electronics is designed to operate in this range.
    Characteristics HD9007
    Luran S777K ( BASF) antistatic UV-resistant thermoplastic material with low thermal conductivity and high reflection.
    White power-painted anticorodal aluminium support bracket. Stainless steel U-bar mounting bracket for shafts from 25 to 44mm.
    Dimensions: external Ø : 130 mm.
    Height, excluding bracket: HD9007 A1: 190 mm, weight: 640 gr.
    HD9007 A2: 240 mm, weight: 760 gr.
    Sensor fixing ring nuts: Ø 27 mm, Ø 25 mm on demand when ordering.
    HD9007 ring-shield is suitable to protect temperature and RH/ temperature sensors used in weather stations from solar radiations, rain and wind.

    TECHNICAL DATA HD9008TR HD9009TR
    Electronics Working Temperature -40… + 80° C
    Sensor Working Temperature -40… + 80° C
    Transmitter Power Supply 7… 30Vdc ( 4… 20mA)
    7… 30Vdc ( 2mA)
    Capacity 300 pF typ.
    HUMIDITY
    Measuring Range 5… 98% RH
    Accuracy at 20° C ± 2% RH ( 5… 90% RH)
    ± 2.5% RH ( for the remaining range values)
    Response time at 63% of final variation 3 min. with filter; 6s without filter
    Output Signal 0% RH = 4.0mA
    100% RH = 20.0mA
    0% RH = 0.00 Vdc
    100% RH = 1.00 Vdc ( * )
    Load Resistance RLmax= ( Vdc - 7)
    22mA RinMIN= 10k©
    TEMPERATURE
    Measuring Range – Standard
    Configuration - ( * * ) -40… + 80° C -40… + 80° C
    Accuracy ± 0.1° C ± 0.1% of measurement
    Response time at 63% of final variation 60s with fi lter; 5s without filter
    Output Signal -40° C = 4.0mA
    + 80° C = 20.0mA
    -40° C = 0.00 Vdc
    + 80° C= 1.00 Vdc ( * )
    Load Resistance RLmax= ( Vdc - 7) 22mA RinMIN= 10k©
    Dimensions Ø 26 x 225mm
    Cable Dimensions
    Maximum Length ( * * * ) 200m 10m
    Wire Min. Section 20 AWG - 0.5mm2 20 AWG - 0.5mm2
    Cable Max. Diameter Ø 5mm Ø 5mm