HD37AB17D and HD37B17D are data loggers able to measure and store at the same time the following parameters:
• RH relative Humidity
• T Room Temperature
• CO Carbon Monoxide ( only HD37AB17D)
• CO2 Carbon Dioxide
Technical Features
Dimensions 275 mm x 45 mm x 40 mm
Weight 230 g ( batteries included)
Materials ABS
Mains power supply ( code SWD06) Batteries charger 100-240Vac/ 6Vdc-1A
Batteries Package with 2 rechargeable batteries 1.2V
type AA ( NiMH)
Autonomy 8 hours of continuous working in measurement mode
Current absorbed with 200¼ A
instrument off
Instrument working temperature 5° C ... 50° C
Working relative humidity 5% RH ... 95% RH no condensed
Temperature / Storage humidity -25° C ... + 70° C / 10% RH ... 90% RH
no condensed
Safety of the stored data Unlimited
Connections
USB interface USB 2.0 cable B type Baudrate 460800
Charger Batteries power supply 2 - poles connector ( positive at the centre)
( code SWD06) Output voltage: 6Vdc
Maximum current: 1600mA ( 9, 60 VA Max) .
Measuring rate 1 sample every three seconds
Storage capacity 20000 Records
Every records includes the following:
- date and time
- measurement of the carbon dioxide ( CO2)
- measurement of the carbon monoxide
( CO- only HD37AB17D)
- measurement of the relative humidity ( RH)
- measurement of the temperature ( T)
Air Quality - CO - CO2
Air Quality - CO - CO2
Logging interval selectable within: 3, 6, 12, 15, 30, 60 seconds, 2, 3, 4, 5 minutes.
The stored values represent the average value of the samples that are stored every
three seconds.
Printing interval selectable within: 3, 6, 12, 15, 30, 60 seconds, 2, 3, 4, 5 minutes.
The printed values represent the average value of the samples that are stored every
three seconds.
Sensor Features
Relative Humidity RH Capacitive sensor
Sensor protection Net filter made of stainless steel ( upon request filter P6 in AISI316 sintered 20¼ m or filter P7
in PTFE sintered 10¼ m)
Measurement range 5...100 % RH
Sensor working range -20...+ 60° C
Accuracy ± 2% ( 10÷ 90% RH) ± 2, 5% in the remaining range
Resolution 0, 1%
Thermal effects ± 2% on all the temperature range
Hysteresis and repeatability 1% RH
Temperature dependence ± 2% on all temperature range
Response time ( T90) < 20 sec. ( air speed = 2m/ sec) without filter
Long term stability 1% / year
Temperature T
Sensor type NTC 10K©
Measurement range -20...+ 60° C
Accuracy ± 0.2° C ± 0.15% of the measure
Resolution 0, 1° C
Response time ( T90) < 30 sec. ( air speed = 2m/ sec)
Long term stability 0.1° C/ year
Carbon monoxide CO ( only HD37AB17D)
Sensor Electro chemical cell
Measurement range 0… 500ppm
Sensor working range -5… 50° C
Accuracy ± 3ppm+ 3% of the measure value
Resolution 1ppm
Response time ( T90) 5 years in normal environmental conditions
Carbon dioxide CO2
Sensor NDIR with a double wave length
Measurement range 0… 5000 ppm
Sensor working range -5… 50° C
Accuracy ± 50ppm+ 3% of the measurement
Resolution 1ppm
Thermal effects 0, 1% f.s./ ° C
Response time ( T90) < 120 sec. ( air speed = 2m/ sec)
Long term stability 5% of the measure/ 5 years