    Description
    HD 2047 is a handheld instrument designed to test and calibrate instruments with
    Pt100 input ( 100© a 0° C) and voltage/ current outputs, such as active or passive temperature transmitter, recorders, testers and data loggers, etc.
    HD 2047 simulates up to 24 fixed values, following the Pt100 scale from -100° C up to + 500° C, with a 2, 3 or 4 -wire connections. The set points selection is made through a rotary switch placed on the front part of the instrument. Whatever operating mode is chosen, Pt100 output is always active.
    HD 2047 can measure with high accuracy voltage/ current outputs of any transmitter connected to the instrument input: -20V...+ 20V continuous voltage range and 0...22mA continuous current range.
    The instrument can also calibrate and test the working process of a passive transmitter and simulates the Pt100 temperature input, providing power supply to
    the transmitter and measuring the current inside it with no external power supply required.
    The instrument is provided with three different keys:
    ON/ OFF turns the instrument on and off. Upon power on, HD 2047 sets automatically to voltage measuring mode.
    MODE selects cyclically the operating mode; by pressing the key consecutively, you select in the following order:
    1. voltage measurement;
    2. current measurement;
    3. current measurement with 4...20mA loop power supply.
    RANGE during the measuring process ( voltage/ current) , this mode allows the full scale selection and the best resolution for the measuring in
    progress: -1.999...+ 1.999, -19.99...+ 19.99 and -199.9...+ 199.9.
    L’ HD 2047 is internally protected against any kind of connecting error made by the
    operator: it is highly recommended anyway not to exceed voltage/ current limits
    shown in technical data.
    A battery sign on the display lights up to indicate that batteries are low and must
    be replaced.

    Technical data ( @ 20° C)
    GENERAL
    Power supply
    4 batteries 1.5V, AA size ( the input for the 9Vdc external supplier is provided only upon request)
    Operating time using 1.5V
    batteries with capacity of
    2250mAh 160 h ( in “ V READ” and “ mA READ” operating mode) 30 h @ loop current = 12mA ( in “ 2 WIRE” operating mode)
    Low batteries signal
    The battery sign lights up with a battery voltage of about 3.6V
    Operating temperature -5...+ 50° C
    Operating relative humidity 0...90% R.H. ( not condensing)
    Weight/ dimensions 580 g ( without batteries) / 23x70x230 mm
    CONTINUOUS VOLTAGE MEASURE
    Measure range
    -1.999V...+ 1.999V: 1mV resolution
    -19.99V...+ 19.99V: 10mV resolution
    Accuracy
    ± 1mV: range -1.999V...+ 1.999V
    ± 10mV: range -19.99V...+ 19.99V
    Input resistance 1M©
    Maximum voltage applied to
    terminals 48Vdc
    CONTINUOUS CURRENT MEASURE
    Measure range
    0.00mA...19.99mA: 10¼ A resolution
    0.0...22.0mA: 100¼ A resolution
    Accuracy
    ± ( 0.01mA+ 0.05% of range) : range 0.00mA...19.99mA
    ± 0.1mA: range 0.0mA...22.0mA
    Shunt resistance 20©
    Overload protection Current limit: 25mA
    PASSIVE TRANSMITTERS: POWER SUPPLY/ MEASURE
    Measure range
    0.00mA...19.99mA: 10¼ A resolution
    0.0...22.0mA: 100¼ A resolution
    Accuracy
    ± ( 0.01mA+ 0.05% of range) : range 0.00mA...19.99mA
    ± 0.1mA: range 0.0mA...22.0mA
    Shunt resistance 20©
    Overload protection Current limit: 25mA
    Maximum load @ 20mA 700©
    Applied voltage 14Vdc
    SIMULATING A Pt100
    Type of RTD Pt100 ( 100© a 0° C, ± = 0.003850, EN60751, IEC751, BS1904)
    Temperature values 24 fixed values from -100 up to + 500° C
    Accuracy ± 0.05% of simulated value
    Room temperature effec ± 5ppm / ° C
    Maximum power loss 125mW
    Maximum load current 20mA