  • Katalog Produk

    THERMOCOUPLE THERMOMETERS: K, J, T, N, R, S, B, E, Model : HD 2108.1 - HD 2108.2 HD 2128.1 - HD 2128.2, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    The HD2108.1 and HD2108.2 with one input and the HD2128.1 and HD2128.2
    with two inputs are portable instruments with a large LCD display. They measure
    the temperature using immersion, penetration air or contact probes. The sensor
    may be a thermocouple of type K, J, T, N, R, S, B or E. The HD2108.2 and HD2128.2 instruments are dataloggers. The HD2108.1 memorizes up to 76, 000 samples, the HD2128.2 up to 38, 000 pairs of values. These data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing interval, printing, and
    baud rate can be configured using the menu.
    All the models are fitted with an RS232C serial port and can transfer the acquired
    measurements in real time to a PC or to a portable printer.
    The Max, Min and Avg function calculate the maximum, minimum or average
    values. Other functions include: the relative measurement REL, the HOLD
    function, and the automatic turning off that can also be disabled. The HD2128.1
    and HD2128.2 calculate the A-B difference of the temperatures captured by the
    two input channels. The instruments have IP67 protection degree.

    INSTRUMENT TECHNICAL CHARACTERISTICS
    Instrument
    Dimensions ( Length x Width x Height) 185x90x40mm
    Weight 470g ( complete with batteries)
    Materials ABS, rubber
    Display 2x41D2 digits plus symbols
    Visible area: 52x42mm
    Operating conditions
    Operating temperature -5… 50° C
    Warehouse temperature -25… 65° C
    Working relative humidity 0… 90% RH without condensation
    Protection degree IP67
    Power
    Batteries 4 1.5V type AA batteries
    Autonomy 200 hours with 1800mAh alkaline batteries
    Power absorbed with instrument off 20¼ A
    Mains Output mains adapter 9Vdc / 250mA
    Measuring unit ° C - ° F - ° K - mV - mV* C
    Security of memorized data Unlimited, independent of
    battery charge conditions
    Time
    Date and time Schedule in real time
    Accuracy 1min/ month max departure
    Measured values storage
    Type - model HD2108.2 2000 pages containing 38 samples each
    Total of 76, 000 samples
    Type - model HD2128.2 2000 pages containing 19 samples each
    38, 000 pairs of samples
    Storage interval 1s… 3600s ( 1hour)
    Serial interface RS232C
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 38400 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Serial cable length Max 15m
    Immediate print interval 1s… 3600s ( 1hour)
    USB interface - model HD2108.2 and HD2128.2
    Type 1.1 - 2.0 electrically isolated
    Connections
    Input for probes 2-pole female polarized standard
    miniature connector
    Serial interface and USB 8-pole MiniDin connector
    Mains adapter 2-pole connector ( positive at centre)
    Measurement of temperature by Instrument
    TC measurement range: K -200… + 1370° C
    TC measurement range: J -100… + 750° C
    TC measurement range: T -200… + 400° C
    TC measurement range: N -200… + 1300° C
    TC measurement range: R + 200… + 1480° C
    TC measurement range: S + 200… + 1480° C
    TC measurement range: B + 200… + 1800° C
    TC measurement range: E -200… + 750° C
    Resolution 0.05° C up to 199.95° C
    0.1° C from 200.0° C to the full scale
    Instrument precision
    Thermocouple K ± 0.1° C up to 600° C
    ± 0.2° C over 600° C
    Thermocouple J ± 0.05° C up to 400° C
    ± 0.1° C over 400° C
    Thermocouple T ± 0.1° C
    Thermocouple N ± 0.1° C up to 600° C
    ± 0.2° C over 600° C
    Thermocouple R ± 0.25° C
    Thermocouple S ± 0.3° C
    Thermocouple B ± 0.35° C
    Thermocouple E ± 0.1° C up to 300° C
    ± 0.15° C over 300° C
    The accuracy only refers to the instrument. Error due to the thermocouple
    or to the cold junction reference sensor is not included.
    Temperature drift @ 20° C 0.02% / ° C
    Drift after 1 year 0.1° C/ year