APPLICATIONS
For continuous analysis of up to 100% hydrogen ( H2) in process gas streams. Also recommended for other applications such as H2 cooled generators, electrolysis units, refinery gas, heat treating atmospheres, blast furnace gas, and many more.
FEATURES
⢠Best value continuous H2 analyzer
⢠Bright digital readout, 4 - 20 mA outputs
⢠Long life thermal conductivity ( TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
⢠Fast response ( T90 - 10 to 15 seconds)
⢠Easy-to-maintain modular layout
⢠Built in sample pump or pressure regulator
⢠Sensor mounted in temperature controlled heated box to improve accuracy
OPTIONS
⢠Hi/ Low gas, low flow, and diagnostic alarms available
⢠Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
⢠Sample conditioning systems available for removal of moisture, acidic gas, and dust
⢠Cabinet purge system available for use in hazardous areas
⢠One-step ' cal-now' button with touch screen LCD display
⢠Full automatic calibration with touch screen LCD display
⢠Cold weather package for operation to -5° F ( -20° C)
⢠Cabinet coolers can be fitted to most models
⢠High temperature filters and probes
CALIBRATION
⢠Nitrogen or air for zero
⢠Span using a known H2 in nitrogen
calibration gas.
DESCRIPTION
The Nova 430 Series Process Hydrogen analyzer utilizes a temperature-controlled TC cell for the detection of H2 in process gas. This cell produces a small millivoltage output which is directly proportional to the H2 detected. This output is then amplified and displayed on a digital readout meter and is also available as an output recorder. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.
MODELS
There are seven types of mounting configurations available:
⢠430N4 - Wall mounted NEMA4 ( IP65) enclosure rating
⢠430N4X - Wall mounted corrosion-resistant NEMA4X
( IP65) enclosure rating
⢠430N12 - Wall mounted NEMA12 ( IP52) enclosure rating
⢠430PM - Panel mounted, slides open for easy access
⢠430RM - 19" ( 483mm) rack mounted, on sliding rails
⢠430N7MC: Wall mounted NEMA7 UL/ CSA explosion-proof with non-intrusive magnetic calibration, meets Class 1 Div 1 Group BCD rating
⢠430PMN7: Wall mounted NEMA7 sensor housing with panel
SPECIFICATIONS
Method of Detection : Temperature-controlled TC cell. Expected life in excess of 10 years. Cell cannot be
burned out due to loss of sample flow.
Ranges Available : Any range between 0-2% and 0-100% H2
Resolution : 0.1% H2 ( 0.01% H2 available)
Accuracy and Repeatability : ± 1% of full scale
Drift : Less than 1% of full scale per month
Response Time ( T-90) : 10-15 seconds to 90% step change
Ambient Temperature Range : 32-122° F ( 0-50° C) . Lower temperatures ( -5° F, -20° C) with Cold Weather Package. Dew point should be at least 9° F ( 5° C) below lowest sample temperature.
Linearity : ± 1% of full scale
Size and Weight : Dimensions will vary depending on enclosure style and options required
Power : 115VAC 60Hz ( 220VAC 50Hz available)
Output Options : 4-20ma into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, MODBUS, Ethernet outputs optional
Alarms : High and/ or low alarm contacts available, relay contacts SPDT 5A @ 220VAC rating.